Since Feb. 2000, DLATR6 (CFD & pre-amp.) was poor condition, because
the discharge broke electric circuit board of DLA.
For that reason, the YSTOP signal from amplifier was lower than others.
Additionally the other board was discharged. (;o;)
Fig 1. is a 2d-image of poor resolution.
The resolution in the x-direction is
poor than y-direction, and we can see some stripes in the x-direction.
We found the time-sum was bad, i.e. another peak was found
in the time-sum spectra.
So we exchanged the broken circuit boards, then got better resolution
than before.(Fig. 2)
As you seen, the resolution in the x-direction of 2d-image 2 was better than
2d-image 1. But some stripes in the x-direction remained.
We checked the time-sum was bad yet and exchanged boards(A1/A2 and A3/A4).
Fig. 3 is a fixed 2d-spectra. But you can see the some spots.
Is it by sputtering?